1 / 5
Medovia Medical
| Model NO. | BK-6310 | Certification | CE, ISO13485 |
| Throughput | 60tests/Hour | Group | Middle-aged and Old |
| Assay Items | 5 Parts, 28 Parameters, and So on | Test Mode | Cbc+5diff, Cbc |
| Measuring Principle | Laser Scatter Techno | Storage | 100,000 Results Storage |
| Lis Interface | Support Hl7 Protocol | Incubation | Integral Titanium Incubation |
| Package Size | 712*587*730mm(W*D*H) | Gross Weight | 104kg |
5-Part Auto Hematology Analyzer
| Measuring Principle | Laser Scatter Technology |
| Sample Modes | Autoloader: 50 tubes, auto mixing, random direction tube barcode scanning technology; Close tube: STAT priority, support both whole blood and capillary blood samples |
| Testing Items | Regular Parameters: WBC, RBC, HGB, HCT, MCV, MCH, MCHC, RDW-SD, RDW-CV, PLT, PDW, MPV, PCT, P-LCR, NEU#, NEU%, LYM#, LYM%, MON#, EOS#, EOS%, BAS#, BAS% Research Parameters: ALY#, ALY%, IG#, IG% |
| Item | Linear range | Carry over | CV |
| WBC | 1.0~99.9x10^9/L | ≤0.5% | ≤2.0% |
| RBC | 0.3~7.0x10^12/L | ≤0.5% | ≤1.5% |
| HGB | 20~240g/L | ≤0.5% | ≤1.5% |
| PLT | 20~999x10^9/L | ≤1.0% | ≤4.0% |